MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO VILLALBA CAPSURE Z NOVUS LEAD; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO VILLALBA CAPSURE Z NOVUS LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 5554-45
Medical Device Problem Codes Signal Artifact/Noise (1036); Electrical /Electronic Property Problem (1198); Over-Sensing (1438); Low impedance (2285)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Continuation of d10: 5054-52 lead was implanted on (b)(6) 2009.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the right atrial (ra) lead exhibited chronic bipolar and unipolar low out of range impedance warnings along with noise and oversensing on stored episodes resulting in false episode detections.Atrial lead position check failure was observed as well.It was also reported that the right ventricular (rv) lead exhibited chronic high thresholds.Both the rv and ra leads remain in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE Z NOVUS LEAD
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key25226778
Report Number2649622-2026-14713
Device Sequence Number18660464
Product Code DTB
UDI-Device Identifier00681490135047
UDI-Public(01)00681490135047(17)110220
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P850089
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/20/2011
Device Model Number5554-45
Device Catalogue Number5554-45
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2026
Initial Report FDA Received Date05/20/2026
Date Device Manufactured03/31/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
W1DR01 IPG
Patient Age38 YR
Patient SexFemale
-
-