The complaint concern bent needle through the hole which caused accidental needle stick injury of nurse.No information about performed test or health consequences were provided.Htl-strefa made several attempts to obtain additional information to the event.The evaluation of archival, customers samples and device history records (dhr) review did not confirm any defects.On the provided photo we can see bent needle from the user end.The needle is exposed from safety pen needle.In the instruction for use there is below information in point 7.Insert the needle into the flat skin at a 90° angle in one continuous movement until the shield is fully recessed.Deliver the dose.Do not remove the needle until the dose has been completely delivered.Count to 10 before removing the needle from the skin to prevent drug leakage.Do not change the direction of the safety pen needle while it remains in the body as this can result in bending or breaking of the needle.We are aware that, the accidental injury might have led or might lead to the accidental blood exposure (abe).It is defined as an accident associated with exposure to blood, bloody fluids or other fluids, and might have caused or might cause serious infection in the injured third party.Due to fact that the event could potentially cause harm and taking into account that health care professional experience difficulties with using the device, we conclude this event should being reported.The final recommendation of hhe team is: to report the event in us where the event had occurred.
|