Additional information: h6.Correction: it was determined after receiving further clarification, this event for this product was not a reportable event per cfr 21, part 803.This supplemental is being filed to identify an mdr was not required for this event.The reported event could not be confirmed.Surgical exploration confirmed that the tmj implants and screws were stable and well fixated, with no evidence of loosening, device malfunction, or limitation in joint function.Medical review concluded that the reported symptoms were not related to the tmj device components.Based on the investigation, there is no indication for an incorrectly working product or any design, material, or manufacturing related issue.Therefore, no corrective and/or preventive actions are deemed necessary at this time.
|