Device evaluation details: the device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Following j&j medtech procedures, a visual inspection and force test of the returned device were performed.Visual inspection reddish material inside the pebax.The device was connected to the carto 3 system and no errors were observed.The force feature of the device was evaluated and the force values and the vector were observed within specifications.No force issues were observed.In a closer inspection observed a hole in the surface of the pebax.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The reddish material inside the pebax could be related to the force issue reported by the customer; therefore, the customer complaint was confirmed.The potential cause of the pebax damage could be related to the manipulation of the device during the procedure; however, this cannot be conclusively determined.The instructions for use (ifu) contain the following recommendations: do not scrub or twist the tip electrode because damage to the tip electrode bond may occur and loosen the tip electrode, or damage the contact force sensor and affect measurement accuracy.In addition, in order to prevent damage to the catheter tip, verify compatibility between the sheath and catheter, advance the catheter through the sheath prior to insertion.Any sheath < 8.5 f is contraindicated.As part of johnson & johnson medtech¿s quality process, all devices are manufactured, inspected, and released to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.# (b)(4).
|