| Model Number |
2110 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: the device was received for evaluation.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.Visual and functional tests were performed, which found a ripped downstream occlusion (dso) seal.The dso seal was replaced to fix the issue.
|
| |
|
Event or Problem Description
|
|
The device was returned because the battery was not charging on the unit.The results of the investigation found that the device's downstream occlusion (dso) seal was ripped.There was no patient involvement.
|
| |
|