| Model Number |
URF-P6 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to failure of the adhesive.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The uretero-reno videoscope was returned to the service center with a report of there's black spots in the image think there are broken fibers.This does not indicate an mdr reportable event.However, mdr reportable failure was found during testing at the service center.During inspection of the device, chipped adhesive of the bending rubber was found.There was no patient involvement.
|
| |
|