The device was returned to olympus for inspection.Reasonably known information suggests probable causes such as component failure, expected or random component failure without any design or manufacturing issue is due to degradation problem identified, problems that occur when the device becomes worn, weakened, corroded, or broken down due to processes such as aging, permeation, and corrosion.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
|
The cysto-nephro videoscope was returned to the service center with a report of leakage.This does not indicate an mdr reportable event.However, mdr reportable failure was found during testing at the service center.During inspection of the device, corroded objective lens was found.There was no patient involvement.
|