Catalog Number 47177E
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Three sets were returned for investigation.The set was examined for defects and abnormalities.No defects or abnormalities were observed.The sets were primed with water and pressurized at the mal luer connection point.No observation of leak.The customer complaint was unable to be replicated.The root cause could not be determined because the issue could not be replicated.A device history record review was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.
 
Event or Problem Description
It was reported that bd alaris smartsite gravity set had connection issues.The following information was received by the initial reporter with the following verbatim.It was reported by the customer that the primary tubing is not making a complete seal at the needle catheter hub and allowing the normal saline to leak out and not flow into the client.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS SMARTSITE GRAVITY SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key25233392
Report Number9616066-2026-01335
Device Sequence Number20087969
Product Code FPA
UDI-Device Identifier10885403235436
UDI-Public(01)10885403235436(11)251215(17)281214(10)25125443
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K022209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number47177E
Device Lot Number25125443
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2026
Initial Report FDA Received Date05/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-