OLYMPUS WINTER & IBE GMBH TELESCOPE "OES ELITE", 4 MM, 30°, HD, AUTOCLAVABLE; RIGID ENDOSCOPE TELESCOPE
|
Back to Search Results |
|
| Model Number |
WA2T430A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the chipped lens in the optical system was traced to component failure.The date of event is unknown.Additional information will be provided if available.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the rigid endoscope telescope exhibited a chipped lens in the optical system.There was no patient involvement.
|
| |
|