| Model Number |
1845000 |
| Medical Device Problem Codes |
Overheating of Device (1437); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3 - evaluation of the returned device was not completed at the time of this report.A follow up report will be sent when analysis is complete.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that pre-op, the indigo drill overheats, the overheating occurred less than a minute of use and the device was running at maximum speed in forward mode and no irrigation was used and the device does not retail the tool (does not stick the tiller).There was no patient involvement.
|
| |
|