| Model Number |
471344-17 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The curved bipolar dissector instrument was analyzed and found to have damaged conductor wire insulation at the yaw pulley.There was fragmentation of the insulation, and the internal conductor wires were exposed.Although the conductor wire insulation was damaged, the internal wire was not frayed or fully broken.The instrument passed the electrical continuity test.Further inspection found no abnormally sharp or damaged components that could have contributed to the cable breaking.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
|
| |
|