MAUDE Adverse Event Report: IRHYTHM TECHNOLOGIES, INC ZIO AT; DETECTOR AND ALARM, ARRHYTHMIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRHYTHM TECHNOLOGIES, INC ZIO AT; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Skin Inflammation/ Irritation (4545)
Date of Event 04/27/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the complaint revealed that the patient wore the zio at for 2 of the 7-day prescribed wear period.The patient does have sensitive skin and does have history of reactions to medical adhesive.While the cause of the reported event cannot be determined, the patient¿s preexisting allergy cannot be ruled out as a contributory factor.Skin irritation and allergic reaction is a known inherent risk of the device.The device manual warnings section read as follows: do not use the zio at patch on patients with known allergic reaction to adhesives or hydrogels or with family history of adhesive skin allergies.Patient may experience skin irritation.If skin irritation such as severe redness, itching or allergic symptoms develop, remove the zio at patch from the patient¿s chest.This event is being reported per 21 cfr 803 as a serious injury.This report does not constitute an admission by rhythm that the product described in this report has any defects or has malfunctioned.These terms are included in form fda 3500a and are fixed terms for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.
 
Event or Problem Description
The presented to urgent care with skin irritation allegedly caused by the zio at.The patient reported experiencing a spreading rash, itching and hives.The patient stated that they removed the device and observed bleeding at the application site.The patient was evaluated by a healthcare provider (hcp) at urgent care, where they received a steroid injection and were prescribed topical cortisone cream for treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIO AT
Common Device Name
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
IRHYTHM TECHNOLOGIES, INC
699 8th st suite 600
san francisco CA 94103
Manufacturer (Section G)
IRHYTHM TECHNOLOGIES, INC
6550 katella avenue, suite 200
cypress CA 90630
Manufacturer Contact
tara mackinnon
699 8th st suite 600
san francisco, CA 94103
6094600577
MDR Report Key25240504
Report Number3007208829-2026-00186
Device Sequence Number21466194
Product Code QYX
UDI-Device Identifier00869770000210
UDI-Public(01)00869770000210(17)261011
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K240117
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/05/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/03/2026
Initial Report FDA Received Date05/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/14/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexMale
-
-