MAUDE Adverse Event Report: MICROPORT CRM S.R.L. CPR4; PACEMAKER/ICD/CRT NON-IMPLANTED COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT CRM S.R.L. CPR4; PACEMAKER/ICD/CRT NON-IMPLANTED COMPONENTS Back to Search Results
Model Number CPR4 HEAD
Medical Device Problem Code Telemetry Discrepancy (1629)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/24/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Since the subject device has not been returned yet, no investigation could be performed.Results will be provided on the next report.
 
Event or Problem Description
Reportedly, on (b)(6) 2026, it was impossible to interrogate an alizea implant.Cpr4 head led did not light on.Interrogation was retried with another alizea implant and it did not work, it was also retried with an eno implant and it worked.Another programmer was used to interrogate alizea implant and it worked normally.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CPR4
Common Device Name
PACEMAKER/ICD/CRT NON-IMPLANTED COMPONENTS
Manufacturer (Section D)
MICROPORT CRM S.R.L.
via crescentino s.n
saluggia (vc) 13040
IT  13040
Manufacturer (Section G)
MICROPORT CRM S.R.L.
via crescentino s.n
saluggia (vc) 13040
IT   13040
Manufacturer Contact
elodie vincent
via crescentino s.n
saluggia (vc) 13040
IT   13040
146013665
MDR Report Key25240872
Report Number1000165971-2026-00279
Device Sequence Number9563270
Product Code OSR
UDI-Device Identifier08031527019611
UDI-Public(01)08031527019611
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 05/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCPR4 HEAD
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date05/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-