| Catalog Number |
HERO® |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Thrombosis/Thrombus (4440); Unintended Loss of Vascular Access (4867)
|
| Date of Event |
04/22/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is not expected to return for evaluation.A follow-up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The subject in a clinical trial experienced clotted access at dialysis which alleges deficiencies related to the durability, reliability, safety, effectiveness, and performance of the device.Patient referred to access center for declot.Patient had successful thrombectomy and angioplasty on (b)(6) 2026.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and three additional relevant complaints for this lot number were found.The product history record was reviewed, and no exception documents were found.Nonconformances of this nature are monitored by the operation team.This complaint will be used to trend for similar complaints.
|
| |
|
Search Alerts/Recalls
|