Catalog Number SXPP1B432
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #(b)(4) date sent to the fda: 5/21/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, the following was obtained: ¿ how many devices demonstrated the alleged deficiency? one and it was left in the patient.The suture split in 2 and barb¿s were on each side.¿ were the suture piece(s) retrieved during the same procedure? no ¿ were x-rays taken to locate the suture piece(s)? no ¿ what measures were taken to retrieve the broken pieces? none ¿ was there any additional tissue damage as a result of searching for the suture piece? n/a ¿ if not retrieved, in what tissue structure the suture was retained? is there any plan in place to remove the suture in the future? please provide details.No as it¿s absorbable suture.¿ please provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep) dr (b)(6) (please refer to the attached email), general surgeon.He wanted us to send in the rest of the box for analysis.¿ please perform and document the follow up attempt for product return.Please provide tracking number- waiting for return box/label to date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.It was reported the sales rep that there are 10 each that need to be returned.The sutures split in half in the patient and was left in the patient.The number of items in the script would only allow the operator to enter a single digit number but there are 10 each to be returned.There is no adverse impact on patient/user reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 5/21/2026.Corrected information: b1, h1 - additional information was received that this event no longer meets malfunction reporting criteria.Therefore this medwatch report is no longer reportable.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATAFIX
Common Device Name
SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo PR 00754
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key25246804
Report Number2210968-2026-05671
Device Sequence Number18778522
Product Code NEW
UDI-Device Identifier10705031246539
UDI-Public(01)10705031246539(17)270430(10)108RBD
Combination Product (Y/N)Y
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K150670
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSXPP1B432
Device Lot Number108RBD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2026
Supplement Date Received by Manufacturer05/21/2026
Initial Report FDA Received Date05/21/2026
Supplement Report FDA Received Date05/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/31/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-