• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC. THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S; ULTRASONIC SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC. THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S; ULTRASONIC SURGICAL DEVICE Back to Search Results
Model Number TB-0535FCS
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/16/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection when the reportable malfunction was observed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the observed malfunction was attributed to degradation problems, however, a definitive root cause could not be established.The event can be detected/prevented by following the applicable chapters in the instructions for use which state: 17 operation 17.1 warnings and cautions page 16.Should any irregularity (error window, abnormal noise, abnormal output, abnormal operation, abnormal appearance, etc.) or malfunction be observed while using the thunderbeat, stop the use and withdraw the instruments from the body cavity.Do not withdraw the transducer plug from the ultrasonic generator.Check the equipment by referring to chapter 8, ¿troubleshooting¿, in the instruction manual for the ultrasonic generator.If the irregularity remains after troubleshooting, replace the transducer.If this does not resolve the issue, replace the thunderbeat instrument.If the irregularity remains unresolved, contact olympus.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was observed during olympus' device inspection that the thunderbeat surgical instrument's grasping part insulation pad was detached.There were no reports of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S
Common Device Name
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key25250177
Report Number3011050570-2026-00569
Device Sequence Number18775194
Product Code GEI
UDI-Device Identifier04953170409677
UDI-Public04953170409677
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K211838
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Notification
Type of Report Initial
Report Date (Section B) 05/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberTB-0535FCS
Device Lot NumberKR489181
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/12/2026
Initial Report FDA Received Date05/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/30/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3003995201-12/09/2024-00
Patient Sequence Number1
Patient SexUnknown
-
-