• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DJO SURGICAL; GLENOID, HEAD W/RETAINING SCREW, RSP, 32MM/NEUTRAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DJO SURGICAL; GLENOID, HEAD W/RETAINING SCREW, RSP, 32MM/NEUTRAL Back to Search Results
Catalog Number 508-32-101
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/30/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The agent reported, "the surgeon decided to reposition the base plate and downsize the glenoid head to a 32-4.Age/g: unk." the previous surgery and the surgery detailed in this event occurred 6 days apart.Item number: 508-32-101 "item description: ar, small socket insert, 32mm neutral eplus " lot#: 862c5136 part revision:h product type: shoulder manufacture date: 30-sep-2025 expiration date: 22-sep-2031.This evaluation is limited in scope as the item(s) associated with this investigation was not returned to djo surgical - austin for examination.The surgery was completed as intended.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate that the reported device(s) was defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.Root cause: the root cause of this complaint was a revision surgery due to the surgeon's decision to reposition the base plate and downsize the glenoid head to 32-4.No information was submitted with the complaint regarding the patient's pre-existing conditions or any patient activities that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of enovis surgical.Containment: inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.Device history records review a review of the device history record(s) shows that the reported component(s) used in the previous surgery met design and manufacturing requirements at the time of release for use.There were no nonconforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Complaint history: customer complaint history of the reported device(s) showed no present trends or ongoing issues that need review.
 
Event or Problem Description
Revision surgery - the surgeon decided to reposition the base plate and downsize the glenoid head to a 32-4, unknown reason.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
GLENOID, HEAD W/RETAINING SCREW, RSP, 32MM/NEUTRAL
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd
austin, TX 78758
5128346313
MDR Report Key25250362
Report Number1644408-2026-00981
Device Sequence Number18778435
Product Code PHX
UDI-Device Identifier00888912024730
UDI-Public00888912024730
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K051075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number508-32-101
Device Lot Number862C5136
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2026
Initial Report FDA Received Date05/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/30/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
506-04-130 LOT: 5039A1078.; 506-04-130 LOT: 5039A1080.; 506-04-134 LOT: 5041A1040.; 509-02-032 LOT: 951W3109.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-