| Model Number |
V*HOME, ENGLISH |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H6: the device was evaluated by ventec where the reported issue of its exhaled tidal volume (vte) levels being low was confirmed.Ventec replaced the control board to resolve the reported issue.Proper device operation was then confirmed through functional and performance testing.The investigation determined that the cause of the reported issue was the control board.
|
| |
|
Event or Problem Description
|
|
It was reported that the device needed service.During the device evaluation, ventec observed that its exhaled tidal volume (vte) levels were low.There were no reports of patient use associated with the reported issue.
|
| |
|
Search Alerts/Recalls
|