U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ENCOR ENSPIRE SYSTEM; REUSABLE BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ENCOR ENSPIRE SYSTEM; REUSABLE BIOPSY Back to Search Results
Catalog Number E4115
Medical Device Problem Code Application Program Problem (2880)
Health Effect - Clinical Code Hematoma (1884)
Date of Event 04/24/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the encor foot pedal vacuum button was not working during two mammogram procedures.During the first procedure, the user forgot that the encor driver also had the vacuum button, so the user skipping the vacuuming after sampling which caused a 2.3 cm hematoma.During the second procedure, after they saw another hematoma for the second patient, then they used the vacuum button on the encor driver and was able to remove the additional fluid in the breast.Because of this the second patient hematoma was not as significant.As the enspire sn: (b)(6) was used during both procedures, and it is unknown whether the foot pedal or the enspire caused the issue.It was identified that encor enspire sn: (b)(6) and encor foot pedal sn: (b)(6) were both used together as a system during both events listed above.There was patient injury during the two events since both patients were presented with hematomas.However, the current status of the patient is unknown.
 
Additional Manufacturer Narrative
H11: as the serial number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCOR ENSPIRE SYSTEM
Common Device Name
REUSABLE BIOPSY
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key25251666
Report Number2020394-2026-01301
Device Sequence Number13387190
Product Code KNW
UDI-Device Identifier00801741086113
UDI-Public(01)00801741086113(17)451130
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
K111100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberE4115
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2026
Initial Report FDA Received Date05/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
Patient SexFemale
Patient Weight59 KG
-
-