• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA GMK PRIMARY TOTAL KNEE SYSTEM; TIBIAL INSERT PS FIX S.5 / 10 MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA GMK PRIMARY TOTAL KNEE SYSTEM; TIBIAL INSERT PS FIX S.5 / 10 MM Back to Search Results
Model Number 02.07.0510PSF
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Medical device site infection (4845)
Date of Event 05/05/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 22 may 2026: lot.135571: (b)(4) items manufactured and released on 10-feb-2014.Expiration date: 2018-dec-31.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Root cause: infection is a known possible complication of any surgery.Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.
 
Event or Problem Description
Revision surgery due to infection after about 11 years and 4 months from primary.The surgeon performed a washout and revised the p.S.10mm s5 insert to a same size insert.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK PRIMARY TOTAL KNEE SYSTEM
Common Device Name
TIBIAL INSERT PS FIX S.5 / 10 MM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro 6874
SZ  6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro 6874
SZ   6874
MDR Report Key25253482
Report Number3005180920-2026-00461
Device Sequence Number10545179
Product Code JWH
UDI-Device Identifier07630030818226
UDI-Public07630030818226
Combination Product (Y/N)N
PMA/510(K) Number
K090988
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2018
Device Model Number02.07.0510PSF
Device Lot Number135571
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2026
Initial Report FDA Received Date05/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/10/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-