Product analysis: analysis was unable to confirm the customer comment of the external pulse generator (epg) randomly powering off with new batteries i nstalled, as the epg passed all automated and bench testing with no anomalies observed.As a preventative measure, the battery drawer shorting bar was replaced.Additionally, the hanger was found broken, and one case screw was contaminated.All found defective parts were replaced and all other identified issues were resolved.The instrument then passed all final functional tests.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|