| Model Number |
CA1032800 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Mechanical Problem (1384); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported failure of ventilator inoperative is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.The device mentioned in this complaint has exceeded its useful life per the applicable ifu.
|
| |
|
Event or Problem Description
|
|
The trilogy 200 ventilator was returned to the manufacturer service center for remediation.The device was not in use on a patient at the time of the occurrence.During the evaluation the device the technician confirmed a ventilator inoperative condition occurred.In conclusion the internal battery and blower assembly were replaced to resolve the issue.The device was retested and past.Additionally, during the service of the device, components were replaced due to cosmetic/physical damage or as part of maintenance of the device.The software was upgraded unrelated to the reportable malfunction/issue.This device been serviced, inspected, tested, met acceptance criteria in accordance with all the applicable customer requirements as defined in the contractual agreement with plexus.The device history record has been reviewed and approved by plexus quality assurance.
|
| |
|
Search Alerts/Recalls
|