• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CONCEAL; PROSTHESIS, PENIS, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CONCEAL; PROSTHESIS, PENIS, INFLATABLE Back to Search Results
Model Number 72018201
Medical Device Problem Codes Mechanical Problem (1384); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/16/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Model number/catalog number: 72404310.Serial number: n/a.Batch/lot number: 1100783557.Model/catalog description: pump ms.Unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Mechanical issue and hole are acknowledged within the instructions for use (ifu) and is not related to off label use or failure to follow instructions.Based on the information available, the cause that contributed to the reported event cannot be established; therefore, a conclusion code of cause not established was assigned to this investigation.
 
Event or Problem Description
It was reported that the patient with this inflatable penile prosthesis (ipp) presented to the hospital with a non-functional device.Two weeks later, a surgical procedure was performed and upon inspection, there was a crack found in the reservoir connecting tube.As a result, the physician removed and replaced the reservoir and pump.The patient was stable following the procedure, and there were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONCEAL
Common Device Name
PROSTHESIS, PENIS, INFLATABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
5086521217
MDR Report Key25256570
Report Number2124215-2026-27666
Device Sequence Number20188308
Product Code JCW
UDI-Device Identifier00878953005676
UDI-Public(01)00878953005676(17)220621(10)178273002
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date06/21/2022
Device Model Number72018201
Device Catalogue Number72018201
Device Lot Number0178273002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2026
Initial Report FDA Received Date05/22/2026
Date Device Manufactured06/21/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age68 YR
Patient SexMale
-
-