• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US EXPRESSEW; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US EXPRESSEW; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 214140
Medical Device Problem Codes Difficult to Open or Close (2921); Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: investigation summary the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found that the device comes in used condition as expected in reusable devices.The jaw was found loose and it is no possible to close or open when pulling the trigger.The device was manufactured on january 2012 which is 14 years old.No other anomalies were noted.A functional test was unable to be performed due to the post-manufacturing damage.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific instructions for use (ifu) for information related to end of life, reprocessing instructions, and inspection procedures.The overall complaint was confirmed as the observed condition of the expressew iii w/o hook would have contributed to the complained device issue.¿ based on the investigation findings, the potential cause is traced to end of life.As per ifu; visually inspect the instrument and check for damage and wear.Jaws and teeth should align properly.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review
=
> a manufacturing record evaluation was performed for the finished device 10260-120109-03 number, and no non-conformances were identified.
 
Event or Problem Description
It was reported that during an unspecified arthroscopic procedure it was observed that the jaw of the expressew iii w/o hook device was not closing correctly and the expressew ii flexible suture passer -ns device had a bent jaw causing the needle to not deploy.Another device was used to complete the procedure.During in-house engineering evaluation it was determined that the jaw was found loose and it is no possible to close or open when pulling the trigger.There was no delay in the procedure reported.There were no adverse patient consequences reported.The exact date of the event was not reported; however, it was reported that the event occurred in 2026.All available information has been disclosed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESSEW
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key25256802
Report Number1221934-2026-02157
Device Sequence Number5417882
Product Code LXH
UDI-Device Identifier10886705020218
UDI-Public(01)10886705020218(10)10260-120109-03
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number214140
Device Lot Number10260-120109-03
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2026
Initial Report FDA Received Date05/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/09/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-