• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 CORAIL; CORAIL IMPLANTS : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 CORAIL; CORAIL IMPLANTS : HIP FEMORAL STEM Back to Search Results
Catalog Number L971211
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Joint Laxity (4526)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2024, the patient underwent bilateral tha.The surgery was completed successfully without any surgical delay.Two years after the surgery, cracks at the stem necks were identified on radiographic imaging.The patient is a ballet dancer and was able to perform ballet movements, including abduction; however, it was reported that a sound suggestive of implant impingement occurred when the hip was brought to maximum abduction.At present, the patient reports no pain or functional issues.No implants have been removed.
 
Additional Manufacturer Narrative
Product complaint: (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: there is no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: d10 concomitant.Corrected: d3, g1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL
Common Device Name
CORAIL IMPLANTS : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex IN 69801
FR  69801
Manufacturer (Section G)
DEPUY FRANCE SAS 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR   69801
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key25256840
Report Number1818910-2026-09152
Device Sequence Number5431923
Product Code MEH
UDI-Device Identifier10603295478621
UDI-Public(01)10603295478621(17)280831(10)4201669
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K213839
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberL971211
Device Lot Number4201669
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/05/2026
Supplement Date Received by Manufacturer05/27/2026
Initial Report FDA Received Date05/22/2026
Supplement Report FDA Received Date05/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORAIL AMT SZ11 STD COL; CORAIL AMT SZ11 STD COL; CORAIL AMT SZ11 STD COL; CORAIL AMT SZ11 STD COL; DELTA CER HEAD 12/14 28MM +1.5; DELTA CER HEAD 12/14 28MM +1.5; DELTA CER HEAD 12/14 28MM +1.5; DELTA CER HEAD 12/14 28MM +1.5
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-