| Catalog Number |
L971211 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Joint Laxity (4526)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024, the patient underwent bilateral tha.The surgery was completed successfully without any surgical delay.Two years after the surgery, cracks at the stem necks were identified on radiographic imaging.The patient is a ballet dancer and was able to perform ballet movements, including abduction; however, it was reported that a sound suggestive of implant impingement occurred when the hip was brought to maximum abduction.At present, the patient reports no pain or functional issues.No implants have been removed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint: (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: there is no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: d10 concomitant.Corrected: d3, g1.
|
| |
|
Search Alerts/Recalls
|