| Model Number |
G2-W-US |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information: the device was not available; therefore, product testing on the actual device could not be performed.The device history record review of the manufacturing lot was performed and there were no non-conformities found to be related to the reported event.A complaint history review was completed.The electronic complaint system was searched for the specified lot#.There was one (1) additional similar issue reported for this lot #.A review of the device labeling including the risk analysis was completed.The reported issue (trocar fragmentation) is identified to be an anticipated hazard.The residual risk for each potential harm was found to be within the insignificant risk region.The ifu adequately provides instructions for stent implantation, precautions, and warnings for the proper use and handling of the device.Mfr# reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during the implantation of the stents from a trabecular microbypass stent system, the trocar broke off in the anterior chamber (ac).Per report, the surgeon completed the procedure, and retrieved the trocar fragment intraoperatively without patient injury.Additional information has been requested.
|
| |
|