Model Number M00505530
Medical Device Problem Codes Break (1069); Adverse Event Without Identified Device or Use Problem (2993); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/29/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Block h6.Device code a150101 is being used to capture the reportable event of cinch failure to deploy.Imdrf code f2301 is being used to capture the reportable event of additional device required.
 
Event or Problem Description
It was reported to boston scientific corporation that two overstitch suture cinches were used during an endoscopic sleeve gastroplasty (esg) procedure on (b)(6) 2026.During the procedure, the cinch failed to deploy due to detachment of the blue slider from the inner wire within the handle.The procedure was completed after manual deployment of the cinch.No patient complications were reported as a result of this event.This is the second of two overstitch suture cinches used in the same procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OverStitch Suture Cinch
Common Device Name
Endoscopic Tissue Approximation Device
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
2546 calle primera
alajuela
CS  
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
+1(914)571-6918
MDR Report Key25257437
Report Number3005099803-2026-02039
Device Sequence Number20188100
Product Code OCW
UDI-Device Identifier00191506060046
UDI-Public(01)00191506060046(17)290302(10)38738005
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00505530
Device Catalogue NumberM00505530
Device Lot Number0038738005
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer05/21/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2026
Supplement Date Received by Manufacturer06/19/2026
07/14/2026
Initial Report FDA Received Date05/22/2026
Supplement Report FDA Received Date07/14/2026
08/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/03/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3005099803-07142026-006R
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-