• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES LLC EDWARDS SAPIEN 3 ULTRA RESILIA TRANSCATHETER HEART VALVE; PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES LLC EDWARDS SAPIEN 3 ULTRA RESILIA TRANSCATHETER HEART VALVE; PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9755RSL29A
Medical Device Problem Codes Migration or Expulsion of Device (1395); Perivalvular Leak (1457)
Health Effect - Clinical Code Hemolysis (1886)
Date of Event 05/02/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Per the instructions for use (ifu), valve migration requiring intervention is a known potential adverse event associated with transcatheter valve replacement (tvr).According to the literature review, valve migration results when forces acting on the transcatheter heart valve (thv) overcome the strength of attachment of the valve to the annulus.Stent valves are subjected to antegrade ejection forces during systole.Less-than-severe and non-uniformly distributed calcification of the leaflets, incorrect bioprosthetic valve sizing, and incomplete frame expansion can contribute to valve migration.Additionally, residual overhanging leaflets can exert downwards force during diastole, causing migration of the thv towards the ventricle.The device training manuals instruct the operator on proper positioning and deployment of the valve, including all procedural and anatomical considerations.Physicians are extensively trained by edwards before they are qualified to use the sapien transcatheter heart valves (all models).Training includes patient screening, device preparation, approach, deployment, imaging, procedure-specific training manuals, and proctored procedures.The correct alignment and positioning of the device at the point of deployment is emphasized as a key factor in the placement and fixation of the device.Operators are also instructed to use fluoroscopy as the primary method of visualization for positioning and deployment.In patients with high-risk anatomical features for migration (i.E., minimal leaflet calcification), balloon valvuloplasty may indicate potential balloon movement during valve deployment.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results suggest that factors indicated above caused or contributed to this event.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.For off-label cases add the following statement prior to the training manuals review paragraph: the device was used in off-label implantation in the mitral position for mac.As there are no specific ifu or training materials related to mitral position for mac procedures, the available training materials were reviewed only for information potentially relevant to the device use.
 
Event or Problem Description
As reported by a field clinical specialist (fcs), 3 days post tavr procedure with a 29 mm sapien 3 ultra resilia (s3ur) valve in the mitral position for mac, another 29 s3ur placed in existing 29 s3ur, as the first s3ur migrated more atrial, significant paravalvular leak (pvl) and hemolysis was noted.No pvl was noted post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 ULTRA RESILIA TRANSCATHETER HEART VALVE
Common Device Name
PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key25259596
Report Number2015691-2026-15471
Device Sequence Number10512093
Product Code NPU
UDI-Device Identifier00690103215823
UDI-Public(01)00690103215823(17)281116(11)251117
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
P140031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number9755RSL29A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2026
Initial Report FDA Received Date05/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-