• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. 4K UHD LCD MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. 4K UHD LCD MONITOR Back to Search Results
Model Number OEV321UH
Medical Device Problem Codes Erratic or Intermittent Display (1182); No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3: date of event is unknown.Additional information will be provided once available.The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
The customer facility¿s biomedical engineer (biomed) reported to olympus, a ¿daisy chained¿ monitor flickers or completely black outs when the video processor or scope is bumped.This monitor is the device that is at the end of the ¿daisy chain¿ and receives its signal from a second monitor.According to the biomed, there is a cable that connects the two monitors together and the cable is routed more than 25 feet and goes under mats on the floor.Although there was no visible damage to the cable, the biomed stated he will order a new cable and replace it to see if it will resolve the problem.This was observed during an unspecified procedure.It was noted that a delay of less than 30 minutes occurred, however, there were no reports of patient or user injury related to this event and the procedure was completed with the same device.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: b5, d9, g6, h2, h6, h11 the device was not returned to olympus for inspection and the reported failure was not confirmed.Based on the results of the investigation, and since the device was not returned for evaluation, the definitive root cause of the reported issue could not be determined.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
No additional information received from the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4K UHD LCD MONITOR
Common Device Name
MONITOR
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key25265167
Report Number3002808148-2026-17780
Device Sequence Number15754447
Product Code GCJ
UDI-Device Identifier04953170418723
UDI-Public04953170418723
Combination Product (Y/N)N
Initial Reporter StateTX
PMA/510(K) Number
K220069
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 07/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOEV321UH
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/23/2026
Supplement Date Received by Manufacturer06/30/2026
Initial Report FDA Received Date05/23/2026
Supplement Report FDA Received Date07/01/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CV-1500.; OEV321UH.; SCOPE.
Patient SexUnknown
-
-