| Model Number |
6640440 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On-site investigation was performed by distributor's field service engineer.The reported issue was reproduced during troubleshooting.According to received information in service report, the replacement of the air gas module solved the issue.The ventilator passed all functional and safety tests and was cleared for clinical use.The air gas module regulates the inspiratory gas flow and a faulty air gas module may affect the delivered volume or gas mixture (o2/air).The device's logs and the claimed part were not provided for further analysis.Our conclusion is that the replaced part was the cause of the failure.
|
| |
|
Event or Problem Description
|
|
It was reported that the ventilator failed internal leakage test and flow transducer test during pre-use check.There was no patient involvement.Manufacturer's ref.:# (b)(4).
|
| |
|