Catalog Number 227402000
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary the product was returned to depuy synthes for evaluation.Visual analysis of the returned device found that the quickset ratchet scdr hdl has the ratcheting mechanism broken.Additionally, device display signs of multiple use for a long period of time.A functional test was performed manually, which revealed that the ratcheting mechanism was not functioning correctly.It is most likely caused due to a breakage in a part of the internal mechanism of the device.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.Additionally, the device was manufactured in 2002, has been in service for over 16 years and shows signs of heavy repeated use.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the quickset ratchet scdr hdl would have contributed to a device issue.Based on the investigation findings, and it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Event or Problem Description
It was reported that the t handle has a crack in it and the screw driver handles collar was damaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RATCHET SCREWDRIVER HANDLE
Common Device Name
HIP INSTRUMENTS : HANDLES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key25277084
Report Number1818910-2026-09181
Device Sequence Number9685556
Product Code HXX
UDI-Device Identifier10603295109020
UDI-Public(01)10603295109020(10)B5KB84000
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number227402000
Device Lot NumberB5KB84000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/15/2026
Initial Report FDA Received Date05/26/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/20/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age75 YR
Patient SexFemale
-
-