Catalog Number PTB8108275
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/14/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may have not been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
It was reported the physician selected a gore® viatorr® tips endoprosthesis with controlled expansion for a transjugular intrahepatic portosystemic shunt procedure.The stent was successfully placed; however, when the stent was approximately 80% deployed, the deployment line broke and the end of the covered portion of the stent remained constrained.Additional manipulation of the introducer sheath and delivery system freed the remaining portion of deployment line and the stent fully deployed.The small piece of deployment line that was still attached to the stent remained in the patient, most likely constrained between the liver and stent.The patient was doing well following the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIATORR® TIPS ENDOPROSTHESIS WITH CONTROLLED EXPANSION
Common Device Name
SHUNT, PORTOSYSTEMIC, ENDOPROSTHESIS
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 1 B/P
32360 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
marci stewart
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key25277876
Report Number3007284313-2026-04805
Device Sequence Number9686026
Product Code MIR
UDI-Device Identifier00733132635047
UDI-Public(01)00733132635047(17)280614
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P040027
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPTB8108275
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/14/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/26/2026
Supplement Report FDA Received Date07/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/16/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age24 YR
Patient SexMale
-
-