• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND EVOLUT FX DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND EVOLUT FX DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number D-EVOLUTFX-2329
Medical Device Problem Codes Unintended Collision (1429); Malposition of Device (2616)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Insufficient Information (4580)
Date of Event 05/08/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Section d information references the main component of the system.Other relevant device(s) are: product id: evfxplus-29, serial/lot #: (b)(6), ubd: 14-nov-2027, udi #: (b)(4).H6.The codes present in section h6.Correspond to multiple components/products that comprise this reported event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that following the successful implant of a transcatheter aortic valve, the paddle housing of the delivery catheter system (dcs) caught on the outflow portion of the valve while withdrawing the dcs, resulting in dislodgement.Subsequently, a non-medtronic valve was implanted in the native annulus.
 
Event or Problem Description
It was reported that following the successful implant of a transcatheter aortic valve, the paddle housing of the delivery catheter system (dcs) caught on the outflow portion of the valve while withdrawing the dcs, resulting in dislodgement.Subsequently, a non-medtronic (edwards sapien) valve was implanted in the native annulus.
 
Additional Manufacturer Narrative
Updated data: d.Suspect medical device: expiration date, lot #, udi # h4.Device mfg date h6.Eval code method pertinent to the dcs was changed from b17 to b18 corrected data - erroneously omitted from the initial 3500a medwatch: h6.Patient code pertinent to the dcs device code pertinent to the valve additional codes: annex g code pertinent to the valve eval code results eval code conclusion.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT FX DCS
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key25278576
Report Number9612164-2026-02705
Device Sequence Number9765854
Product Code NPT
UDI-Device Identifier00763000942700
UDI-Public(01)00763000942700(17)280127(10)0013259444(20)01
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P130021
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberD-EVOLUTFX-2329
Device Catalogue NumberD-EVOLUTFX-2329
Device Lot Number0013259444
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2026
Supplement Date Received by Manufacturer05/27/2026
Initial Report FDA Received Date05/26/2026
Supplement Report FDA Received Date06/24/2026
Date Device Manufactured01/27/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient SexMale
-
-