| Catalog Number |
SRF-295 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
03/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture and capsular contracture, baker grade iii.
|
| |
|
Event or Problem Description
|
|
Patient reported "rupture".Later healthcare professional reported "capsular contracture baker grade iii".This record is for the left side.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation:
per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
rupture: observed broken device through microscopic inspection assessed as surgical damage.No further actions are required as it is not related to the manufacturing process.Capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.No additional observations performed on the device.No further actions are required.Additional, changed, and/or corrected data: a2, b5, b6, d9, h3, h6.
|
| |
|
Event or Problem Description
|
|
Patient reported "rupture".Later healthcare professional reported "capsular contracture baker grade iii".Patient has undergone capsulectomy.This record is for the left side.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, b6, d4, d6b, h6.
|
| |
|
Event or Problem Description
|
|
Device has been explanted and replaced.
|
| |
|