| Catalog Number |
21-6300-51 |
| Medical Device Problem Code |
Failure to Sense (1559)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H6: codes were updated.One device was returned for investigation.It was stated that error code 1535 has occurred on the pump.Checking the logs i can confirm that error code 1535 occurred due to a faulty air sensor.After which, error code 1535 was cleared and the pump passed functional.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.
|
| |
|
Event or Problem Description
|
|
It was reported that the device has an error code lec1535.No patient involvement or adverse event was reported.The event occurred during setup.
|
| |
|
Search Alerts/Recalls
|