MAUDE Adverse Event Report: GYRUS ACMI, INC. THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S; ULTRASONIC SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC. THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S; ULTRASONIC SURGICAL DEVICE Back to Search Results
Model Number TB-0535FCS
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection and the reported malfunction was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the reported event was attributed to degradation related issues, however, a definitive root cause could not be established.The event can be detected/prevented by following the applicable chapters in the instructions for use which state: 17 operation 17.1 warnings and cautions page 16.Should any irregularity (error window, abnormal noise, abnormal output, abnormal operation, abnormal appearance, etc.) or malfunction be observed while using the thunderbeat, stop the use and withdraw the instruments from the body cavity.Do not withdraw the transducer plug from the ultrasonic generator.Check the equipment by referring to chapter 8, ¿troubleshooting¿, in the instruction manual for the ultrasonic generator.If the irregularity remains after troubleshooting, replace the transducer.If this does not resolve the issue, replace the thunderbeat instrument.If the irregularity remains unresolved, contact olympus.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was reported that during therapeutic hysterectomy procedure the thunderbeat's probe tip broke off.The procedure was completed with the same device.There was no patient injury or medical intervention associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S
Common Device Name
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key25279643
Report Number3011050570-2026-00579
Device Sequence Number21224506
Product Code GEI
UDI-Device Identifier04953170409677
UDI-Public04953170409677
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K211838
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTB-0535FCS
Device Lot NumberKR446736 / KR443791-PACKAGE LO
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2026
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/12/2026
Supplement Date Received by Manufacturer05/21/2026
Initial Report FDA Received Date05/26/2026
Supplement Report FDA Received Date06/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-