• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ACCOLADE? SR; Implanted pacemaker

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ACCOLADE? SR; Implanted pacemaker Back to Search Results
Model Number L300
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This product investigation is still in progress.This report will be updated when product investigation is complete.
 
Event or Problem Description
It was reported that this pacemaker recorded a code 1003, indicating the voltage is too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis identified potential high impedance within the battery.Device replacement was recommended.No adverse patient effects were reported.This device remains in service at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE? SR
Common Device Name
Implanted pacemaker
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel
EI  
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key25280098
Report Number2124215-2026-28122
Device Sequence Number9568216
Product Code LWP
UDI-Device Identifier00802526559150
UDI-Public(01)00802526559150(17)190802
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
N970003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 08/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/02/2019
Device Model NumberL300
Device Catalogue NumberL300
Device Lot Number718772
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/26/2026
Supplement Report FDA Received Date09/04/2026
Date Device Manufactured08/18/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2124215-03/18/26-001-C
Patient Sequence Number1
Patient Age82 YR
Patient SexMale
-
-