• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. AUGMENTED REAMER DRIVE SHAFT, LOCKING; ORTHOPEDIC MANUAL SURG INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. AUGMENTED REAMER DRIVE SHAFT, LOCKING; ORTHOPEDIC MANUAL SURG INSTR Back to Search Results
Model Number AUGMENTED REAMER DRIVE SHAFT, LOCKING
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
On 5-may-2026, it was reported by a sales representative via (b)(4) that an ar-9676t drive shaft's locking pin fell off.Issue discovered during a case, device swapped, and case completed with no patient effect.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUGMENTED REAMER DRIVE SHAFT, LOCKING
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key25285454
Report Number1220246-2026-02999
Device Sequence Number10059795
Product Code LXH
UDI-Device Identifier00888867499058
UDI-Public(01)00888867499058(11)250402(10)128141-2
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 07/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAUGMENTED REAMER DRIVE SHAFT, LOCKING
Device Catalogue NumberAR-9676T
Device Lot Number128141-2
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date05/05/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2026
Supplement Date Received by Manufacturer07/01/2026
Initial Report FDA Received Date05/26/2026
Supplement Report FDA Received Date07/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/02/2025
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-