| Model Number |
CYF-5 |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: degradation.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The oes cystonephrofiberscope was returned to the service center with a cloudy image.This does not indicate an mdr reportable event.However, mdr reportable failure was found during testing at the service center.During inspection of the device, a dirty objective lens was found to be most likely due to degradation.There was no patient involvement.
|
| |
|