The agent reported "(revision was needed because the baseplate and the taper disengaged)".The previous surgery and the surgery detailed in this event occurred 1 year and 1 week apart.Item number: 508-65-000 item description: altivate reverse mod baseplate and taper kit, 6.5mm neutral lot number: 5008a1113 part revision: a product type: shoulder manufacture date: 09/25/24 expiration date: 09/17/29.This evaluation is limited in scope as the items associated with this investigation were not returned to djo surgical - austin for examination.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.Root cause: the root cause of this complaint was a revision surgery due to dissociation.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, prolonged overhead activities, inadequate soft tissue support, patient activities or trauma.Containment: inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.Device history records review: a review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was an ncmr# (b)(4) associated with the concomitant part #508-40-101, glenoid, head w/retaining screw, rsp, 40mm, neutral which document that out of 15 parts lot, all items were rejected due to incomplete inspection on is25879.Part #15 missing chamfer.240 max dimension, and part #1 missing radius gage.015+/-.01 dimension.Later, the rejected were reworked and accepted after proper justification.All other items in the lot were met with fit, form and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Complaint history: customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.
|