• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DJO SURGICAL; ALTIVATE REVERSE MOD BASEPLATE AND TAPER KIT, 6.5MM NEUTRAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DJO SURGICAL; ALTIVATE REVERSE MOD BASEPLATE AND TAPER KIT, 6.5MM NEUTRAL Back to Search Results
Catalog Number 508-65-000
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 05/05/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The agent reported "(revision was needed because the baseplate and the taper disengaged)".The previous surgery and the surgery detailed in this event occurred 1 year and 1 week apart.Item number: 508-65-000 item description: altivate reverse mod baseplate and taper kit, 6.5mm neutral lot number: 5008a1113 part revision: a product type: shoulder manufacture date: 09/25/24 expiration date: 09/17/29.This evaluation is limited in scope as the items associated with this investigation were not returned to djo surgical - austin for examination.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.Root cause: the root cause of this complaint was a revision surgery due to dissociation.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, prolonged overhead activities, inadequate soft tissue support, patient activities or trauma.Containment: inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.Device history records review: a review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was an ncmr# (b)(4) associated with the concomitant part #508-40-101, glenoid, head w/retaining screw, rsp, 40mm, neutral which document that out of 15 parts lot, all items were rejected due to incomplete inspection on is25879.Part #15 missing chamfer.240 max dimension, and part #1 missing radius gage.015+/-.01 dimension.Later, the rejected were reworked and accepted after proper justification.All other items in the lot were met with fit, form and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Complaint history: customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.
 
Event or Problem Description
Revision surgery - due to the baseplate and the taper disengaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
ALTIVATE REVERSE MOD BASEPLATE AND TAPER KIT, 6.5MM NEUTRAL
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd.
austin, TX 78758
MDR Report Key25294153
Report Number1644408-2026-00982
Device Sequence Number9319803
Product Code KWS
UDI-Device Identifier00190446822837
UDI-Public00190446822837
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K233481
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number508-65-000
Device Lot Number5008A1113
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2026
Initial Report FDA Received Date05/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/25/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
508-40-101 LOT: 875C1491A; 509-03-040 LOT: 960W1153
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age87 YR
Patient SexMale
-
-