SMITH & NEPHEW, INC. SCREW BIOSURE REGENESORB 8MM X 25MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
72204399 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Perforation (2001); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
05/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a knee arthroscopy, a nitinol guide wire was inserted, the hole was prepared with a 9.0 drill bit and tapped with an 8 mm tap, and a biosure screw was inserted; upon tightening, the screw fractured into multiple pieces.The broken pieces were removed using kelly tweezers.Part of the screw remained in the bone hole.The procedure was resumed, after a delay greater than 30 minutes, using a competitor device in an additionally drilled bone hole.Additional anesthesia was not required.No further complications were reported.The patient's health condition is stable.
|
| |
|
Search Alerts/Recalls
|
|
|