| Model Number |
GIF-H180J |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The gastrointestinal videoscope was returned to olympus with a reported complaint of loose angulation for both control knobs; this does not indicate an mdr reportable event.However, an mdr reportable failure was found during testing at the service center, the light guide lens has dirty, foreign bodies.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.The device evaluation found the light guide lens has dirty, foreign bodies.Based on the results of the investigation, olympus confirmed the device had foreign material, but the type of material or root cause cannot be identified.A definitive root cause cannot be determined.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|