NORIS MEDICAL LTD TUFF PRO IMPLANT; TUFF PRO DENTAL IMPLANT Back to Search Results
Model Number NMAF5016
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/10/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The dhr was reviewed for lot #0019885 and there was no evidence discovered to indicate that a product defect or non-conformity contributed to the issue.To date, items of the same lot have been sold without any similar reported event during the period of review.The investigation for this event has not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Our manufacturing system assures, that production and process controls are in place to ensure that batches conform to the applicable specifications before they are distributed.The implant's expiration date is february 2025, and according to the instructions and labeling, the implant cannot be inserted after this date.It's possible the doctor made a mistake in indicating the year of implantation and removal, as the report date was (b)(6) 2025.However, we don't have an answer from the doctor yet.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2026.On (b)(6) 2026, the implant was not osseointegrated.Loosening of implant not related to bone ingrowth was verified.Don't have information about patient medical history.The device was forwarded to the manufacturer.No patient operative or post-operative complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUFF PRO IMPLANT
Common Device Name
TUFF PRO DENTAL IMPLANT
Manufacturer (Section D)
NORIS MEDICAL LTD
hataasia 8
nesher, 36888 08
IS  3688808
Manufacturer (Section G)
NORIS MEDICAL LTD
hataasia 8
nesher, 36888 08
IS   3688808
Manufacturer Contact
anna dvinska
hataasia 8
nesher, 36888-08
IS   3688808
MDR Report Key25298199
Report Number3011390931-2026-00160
Device Sequence Number9294052
Product Code DZE
UDI-Device Identifier07290108817096
UDI-Public07290108817096
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K140440
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 05/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date02/23/2025
Device Model NumberNMAF5016
Device Catalogue NumberNMAF5016
Device Lot Number0019885
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2026
Initial Report FDA Received Date05/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/24/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age36 YR
Patient SexUnknown
-
-