| Model Number |
1054096 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Failure to Calibrate (2440)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported failure of the active exhalation purge valve and proportional valve test steps is accommodated in the product risk management file as being linked to hazard with description patient exposure to function / deterioration of function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Dhr review: the device mentioned in this complaint has exceeded its useful life per the applicable ifu.
|
| |
|
Event or Problem Description
|
|
A trilogy 100 was returned to the manufacturer's service center for evaluation.There was no harm or injury reported.There was no evidence the device was in patient use.During the evaluation the device failed the active exhalation purge valve and proportional valve test steps.The device's active exhalation control module needs to be replaced to address the issue.The customer refused the repair quote and requested the device be returned unrepaired.The device was returned unrepaired.
|
| |
|
Search Alerts/Recalls
|