Catalog Number 300-01-11
Medical Device Problem Codes Separation Failure (2547); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Pain (1994)
Date of Event 05/06/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient previously underwent a left shoulder replacement and subsequently required a revision surgery to address pain.Based on provided radiograph, the glenoid components are seen to be no longer be fixed within the bone and appear to have rotated.The surgeon¿s plan for the revision procedure was to remove the glenosphere, recut the humerus, and implant new humeral components.During the procedure, the surgeon removed the fixed-angle torque screw and attempted to remove the humeral adapter tray; however, the stem disengaged and was removed with the tray still attached.Attention was then turned to the glenosphere.After removal of the locking screw, the surgeon attempted to disengage the glenosphere from the baseplate using manual manipulation.Additional attempts were made, however, the glenosphere and baseplate were ultimately removed together as a single construct.The surgeon removed all the implants and made a homemade spacer as a first stage revision procedure.The patient was last known to be in stable condition following the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE, HUMERAL STEM PRIMARY, PRESS FIT 11MM
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
katy priest
2320 nw 66 ct
gainesville, FL 32653
3523771140
MDR Report Key25299690
Report Number1038671-2026-00649
Device Sequence Number13087984
Product Code KWS
UDI-Device Identifier10885862079312
UDI-Public10885862079312
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K042021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number300-01-11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2026
Initial Report FDA Received Date05/27/2026
Date Device Manufactured11/26/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization;
Patient Age81 YR
Patient SexMale
-
-