Block h6: imdrf device code a0509 captures the reportable event of suction button sticking.Block h11: at this time the device is currently undergoing technical analysis.The investigation is not yet complete.A supplemental report will be submitted once the results are available.On may 12, 2026, boston scientific initiated a field safety corrective action (fsca - boston scientific reference: (b)(4)) that included the removal of orca? single use air/water and suction valves associated with batch number 38400303 due to increased reports of suction button sticking issues.A communication was sent out to materials managers/health care professionals on (b)(6) 2026, with instructions to immediately stop further use or distribution of orca? single use air/water and suction valves associated with batch number 38400303, remove the devices from inventory, and segregate them in a secure location until they can be returned to boston scientific.Boston scientific will continue to closely monitor and trend similar events and associated risks, as outlined in our quality systems process.
|