| Model Number |
6300 |
| Medical Device Problem Code |
Failure to Infuse (2340)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: the device was received for evaluation.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.The reported issue was verified in the event history log.Visual and functional tests were performed.The customer's problem was duplicated.The cause of the issue was a defective mainboard.The mainboard was replaced to fix the issue.
|
| |
|
Event or Problem Description
|
|
It was reported that the device displayed error code lec1280.The event occurred during setup and there was no patient involvement.
|
| |
|