| Model Number |
CYF-V2R |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: degradation.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The visera cysto-nephro videoscope was returned to the service center with a failed leak test.This does not indicate an mdr reportable event.However, mdr reportable failure was found during testing at the service center.During inspection of the device, a chipped chipped adhesive at the objective lens and light guide lens were found to be most likely due to degradation.There was no patient involvement.
|
| |
|