| Model Number |
ILS-0450-KT |
| Medical Device Problem Codes |
Use of Device Problem (1670); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure, the catheter was not recognized after the first biopsy pass and an out-of-sensing-volume error was observed.An extended working channel (ewc) issue was reported, including inability to deflect the tip.Troubleshooting steps were performed and no magnetic interference was noted, the patient sensor triplet (pst) did not move.The physician was able to complete the electromagnetic navigational bronchoscopy (enb) portion of the case by opening a new catheter kit.There was no patient injury.
|
| |
|