Block h6: imdrf a0509 captures the reportable event of suction valve sticking.Block h7, h9, h11: on may 12, 2026, boston scientific initiated a field safety corrective action (fsca - boston scientific reference: (b)(4)) that included the removal of orca? single use air/water and suction valves associated with batch number 38400303 due to increased reports of suction button sticking issues.A communication was sent out to materials managers/health care professionals on may 12, 2026, with instructions to immediately stop further use or distribution of orca? single use air/water and suction valves associated with batch number 38400303, remove the devices from inventory, and segregate them in a secure location until they can be returned to boston scientific.Boston scientific will continue to closely monitor and trend similar events and associated risks, as outlined in our quality systems process.
|